BH GSO IEC 61223-3-2:2016
IEC 61223-3-2:2007
Bahraini Standard
Current Edition
·
Approved on
08 February 2016
Evaluation and routine testing in medical imaging departments - Part 3-2: Acceptance tests - Imaging performance of mammographic X-ray equipment
BH GSO IEC 61223-3-2:2016 Files
English
null Pages
Current Edition
Reference Language
Obtaining this standard through the store is currently unavailable. You can acquire it directly from its source.
BH GSO IEC 61223-3-2:2016 Scope
This part of IEC 61223 applies to the effectiveness of mammographic X-RAY EQUIPMENT, with
respect to image quality and dose, in combination with aspects of EQUIPMENT safety.
This standard applies to mammographic X-RAY EQUIPMENT and MAMMOGRAPHIC STEREOTACTIC
DEVICES.
The tests described in this standard require the quality and performance of the X-RAY IMAGE
RECEPTORS to be assured prior to the acceptance testing when they are not an integral part of
the mammographic X-RAY EQUIPMENT. This includes RADIOGRAPHIC FILMS, INTENSIFYING
SCREENS, RADIOGRAPHIC CASSETTES, storage phosphor plates and ASSOCIATED EQUIPMENT such
as film processors or storage phosphor plate readers, IMAGE DISPLAY DEVICES and HARD COPY
CAMERAS.
For testing RADIOGRAPHIC CASSETTES and INTENSIFYING SCREENS, this standard makes
reference to ISO 4090. Sensitivity and contrast for the screen-film image receptors are
considered to be stated according to ISO 9236-3.
NOTE Currently there exists no IEC standard for acceptance testing of HARD COPY CAMERAS or IMAGE DISPLAY
DEVICES.
By the measurements described in this standard, data for AVERAGE GLANDULAR DOSE
calculation can be determined.
When the results of the ACCEPTANCE TEST are in compliance with the expected values, the
baseline values for the subsequent CONSTANCY TESTS are established.
This part of IEC 61223 defines
a) the essential parameters which describe the performance of the above-mentioned
mammographic X-RAY EQUIPMENT with regard to image quality and dose; and
b) the methods of testing whether measured quantities related to those parameters comply
with specified tolerances.
These methods mainly rely on non-invasive measurements that use appropriate test
EQUIPMENT and are performed during or after the installation. Signed statements covering
steps in the installation procedure can be used as part of the ACCEPTANCE TEST. Tests
required by a higher level of compliance take precedence over similar tests with a lower level
of compliance. This concept is described in 4.1.
This standard does not in itself specify limiting values or tolerances for the parameters under
investigation.
A difficulty may arise with regard to the responsibility for acceptance testing when the
film/screen combination, film processing chemistry or computed radiography system is
changed. This arises from a combination of causes. Firstly, the image receptor MANUFACTURER
and the X-RAY EQUIPMENT MANUFACTURER may be different. Secondly a change in image
receptor or film processing chemistry may alter the system performance. When system
integration such as the above occurs, it is important that acceptance testing is performed.
When a change occurs which could alter system performance, it is essential that the system
integrator (i.e. whoever is responsible for this change) discusses the implication of their
change with the X-RAY EQUIPMENT MANUFACTURER so that the latter can adjust the imaging
system if necessary.
ACCEPTANCE TESTING of mammographic X-RAY EQUIPMENT requires average skill in medical
physics. However, the decision concerning who performs the test is determined by local rules
(e.g. contract, regulation, law).
Best Sellers From Health Sector
GSO 2017:2023
Gulf Standard
Personal Care and Baby Wet Wipes
BH GSO 2017:2023
GSO 2017:2023
Bahraini Standard
Personal Care and Baby Wet Wipes
BH GSO 2241:2023
GSO 2241:2023
Bahraini Standard
Cosmetic Products - Body Cleaners
GSO 2241:2023
Gulf Standard
Cosmetic Products - Body Cleaners
Recently Published from Health Sector
GSO ASTM F404:2026
ASTM F404:21
Gulf Standard
Standard Consumer Safety Specification for High Chairs
GSO ISO 19223-2:2026
ISO 19223-2:2025
Gulf Standard
Lung ventilators and related equipment — Vocabulary and semantics — Part 2: High frequency and jet ventilation
GSO ASTM F619:2026
ASTM F619:20(2025)
Gulf Standard
Standard Practice for Extraction of Materials Used in Medical Devices
GSO ISO 16687:2026
ISO 16687:2025
Gulf Standard
Impact assessment for museums