BH GSO IEC 61010-2-034:2024
IEC 61010-2-034:2023
Bahraini Standard
Current Edition
·
Approved on
21 November 2024
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-034: Particular requirements for measurement equipment for insulation resistance and test equipment for electric strength
BH GSO IEC 61010-2-034:2024 Files
No files for this standard are available in the store currently.
Obtaining this standard through the store is currently unavailable. You can acquire it directly from its source.
BH GSO IEC 61010-2-034:2024 Scope
IEC 61010-2-034:2023 specifies safety requirements to equipment for measuring insulation resistance and to equipment for testing electric strength which have an output voltage exceeding 50 V a.c. or 120 V d.c. This document also applies to combined measuring equipment which has an insulation resistance measurement function or an electric strength test measurement function. This document has the status of a horizontal publication in accordance with IEC Guide 108. It has the status of a group safety publication in accordance with IEC Guide 104. This second edition cancels and replaces the first edition published in 2017. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
a) in 1.2.1, requirements for protection against hazards which could occur from reading a voltage have been added to the scope;
b) Clause 2, all normative references have been dated; new normative references have been added;
c) in 4.3.2.5, requirements for power supply have been modified;
d) in 4.3.2.6, requirements for inputs/outputs have been modified;
e) in 5.1.5.101.2, minimum ratings for voltage of measuring terminals are required;
f) in 5.4.2, new ratings for documentation have been added;
g) in 5.4.4, new instructions for operation have been added;
h) in 5.101.1, hazard indicators shall be functional in normal condition and in single fault condition;
i) in 6.6.101.1, insulating material of group I may be allowed for determination of creepage distances of measuring circuit terminals;
j) in 6.6.101.2, Clearances and creepage distances above 1 000 V a.c. and 1 500 V d.c. for measuring circuit terminals in unmated position have been defined;
k) in 6.6.101.3, requirements for measuring circuit terminals in partially mated position have been specified;
l) in 6.6.101.4, requirements for measuring circuit terminals in mated position have been specified;
m) Subclause 6.102 replaces 6.9.103 and has been rephrased;
n) new Subclause 9.101 to consider the protection of measuring circuits against the spread of fire and arc flash has been added. Table 102 has been replaced by Table K.101;
o) In 9.101.2, relocation of 101.3 of previous edition;
p) In 9.101.3, relocation of 101.4 of previous edition, extension to measurement category II and reference to IEC 61000-4-5 for tests;
q) in 9.101.4, requirements for measuring circuit terminals in mated position have been specified;
r) in 9.101.5, relocation of K.103 of previous edition with numerous technical changes;
s) in 14.101, relocation of 14.102. 14.101 of previous edition has been removed;
t) in 101.3, relocation of 101.5 of previous edition, and more requirements added against hazard occurring from reading a voltage value;
u) in K.2.1, another method for determination of clearances of secondary circuits is proposed;
v) in K.3.2, new Table K.15 and Table K.16 for clearance calculation;
w) in K.3.101, relocation of 6.9.104 of previous edition;
x) in K.101.4.1, new Table K.103 and Table K.104 replace Table K.102, Table K.103 and Table K.104;
y) in K.101.4, the subclause has been reviewed. Tables and tests for solid insulation have been modified. Table K.105 replaces Table K.9;
z) Table K.101, replacement of Table K.106;
aa) Clause K.4, redraft of the clause to propose a method for determination of Ut for circuits which reduce transient overvoltages;
bb) Annex EE: addition of a new informative annex for determination of clearances for Table 101.
a) in 1.2.1, requirements for protection against hazards which could occur from reading a voltage have been added to the scope;
b) Clause 2, all normative references have been dated; new normative references have been added;
c) in 4.3.2.5, requirements for power supply have been modified;
d) in 4.3.2.6, requirements for inputs/outputs have been modified;
e) in 5.1.5.101.2, minimum ratings for voltage of measuring terminals are required;
f) in 5.4.2, new ratings for documentation have been added;
g) in 5.4.4, new instructions for operation have been added;
h) in 5.101.1, hazard indicators shall be functional in normal condition and in single fault condition;
i) in 6.6.101.1, insulating material of group I may be allowed for determination of creepage distances of measuring circuit terminals;
j) in 6.6.101.2, Clearances and creepage distances above 1 000 V a.c. and 1 500 V d.c. for measuring circuit terminals in unmated position have been defined;
k) in 6.6.101.3, requirements for measuring circuit terminals in partially mated position have been specified;
l) in 6.6.101.4, requirements for measuring circuit terminals in mated position have been specified;
m) Subclause 6.102 replaces 6.9.103 and has been rephrased;
n) new Subclause 9.101 to consider the protection of measuring circuits against the spread of fire and arc flash has been added. Table 102 has been replaced by Table K.101;
o) In 9.101.2, relocation of 101.3 of previous edition;
p) In 9.101.3, relocation of 101.4 of previous edition, extension to measurement category II and reference to IEC 61000-4-5 for tests;
q) in 9.101.4, requirements for measuring circuit terminals in mated position have been specified;
r) in 9.101.5, relocation of K.103 of previous edition with numerous technical changes;
s) in 14.101, relocation of 14.102. 14.101 of previous edition has been removed;
t) in 101.3, relocation of 101.5 of previous edition, and more requirements added against hazard occurring from reading a voltage value;
u) in K.2.1, another method for determination of clearances of secondary circuits is proposed;
v) in K.3.2, new Table K.15 and Table K.16 for clearance calculation;
w) in K.3.101, relocation of 6.9.104 of previous edition;
x) in K.101.4.1, new Table K.103 and Table K.104 replace Table K.102, Table K.103 and Table K.104;
y) in K.101.4, the subclause has been reviewed. Tables and tests for solid insulation have been modified. Table K.105 replaces Table K.9;
z) Table K.101, replacement of Table K.106;
aa) Clause K.4, redraft of the clause to propose a method for determination of Ut for circuits which reduce transient overvoltages;
bb) Annex EE: addition of a new informative annex for determination of clearances for Table 101.
Best Sellers From Electrical Sector
GSO 2530:2016
Gulf Standard
Energy Labelling And Minimum Energy Performance Requirements For Air-Conditioners

GSO 34:2007
Gulf Technical Regulation
LEAD-ACID STARTER BATTERIES USED FOR
MOTOR VEHICLES AND INTERNAL
COMBUSTION ENGINES

BH GSO 34:2015
GSO 34:2007
Bahraini Technical Regulation
LEAD-ACID STARTER BATTERIES USED FOR
MOTOR VEHICLES AND INTERNAL
COMBUSTION ENGINES


GSO 35:2007
Gulf Standard
Methods of test
for lead-acid starter batteries used for motor vehicles
and internal combustion engines

Recently Published from Electrical Sector
GSO 195:1994
Gulf Standard
METHODS OF TEST FOR
ADHESIVE INSULATING TAPES FOR
ELECTRICAL PURPOSES

GSO 31:1984
Gulf Standard
METHODS OF TEST FOR
PVC-INSULATED CABLES AND CORDS
WITH CIRCULAR COPPER CONDUCTORS

GSO 723:1997
Gulf Standard
Methods of test for telephone sets.

GSO 1842:2007
Gulf Standard
METHODS OF TESTING FOR PLUGS AND SOCKET OUTLETS FOR VOLTAGES FROM 127V AND UP TO 250V FOR HOUSEHOLD AND SIMILAR GENERAL USE