GSO ISO 17665-1:2014
ISO 17665-1:2006
Gulf Standard
Current Edition
·
Approved on
25 December 2014
Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
GSO ISO 17665-1:2014 Files
English
38 Pages
Current Edition
Reference Language
80.21 BHD
GSO ISO 17665-1:2014 Scope
1.1 Inclusions
1.1.1 This part of ISO 17665 specifies requirements for the development, validation and routine control of a moist heat sterilization process for medical devices.
NOTE Although the scope of this part of ISO 17665 is limited to medical devices, it specifies requirements and provides guidance that may be applicable to other health care products.
1.1.2 Moist heat sterilization processes covered by this part of ISO 17665 include but are not limited to:
a) saturated steam venting systems;
b) saturated steam active air removal systems;
c) air steam mixtures;
d) water spray;
e) water immersion.
NOTE See also Annex E.
1.2 Exclusions
1.2.1 This part of ISO 17665 does not specify requirements for development, validation, and routine control of a process for inactivating the causative agents of spongiform encephalopathies such as scrapie, bovine spongiform encephalopathy and Creutzfeldt-Jakob disease. Specific recommendations have been produced in particular countries for the processing of materials potentially contaminated with these agents.
NOTE See also ISO 22442-1, ISO 22442-2 and ISO 22442-3.
1.2.2 This part of ISO 17665 does not apply to those sterilization processes that are based on a combination of moist heat with other biocidal agents (e.g. formaldehyde) as the sterilizing agent.
1.2.3 This part of ISO 17665 does not detail a specified requirement for designating a medical device as “sterile.”
NOTE Attention is drawn to national or regional requirements for designating medical devices as “sterile.” See, for example, EN 556-1 or ANSI/AAMI ST67.
1.2.4 This part of ISO 17665 does not specify a quality management system for the control of all stages of production of medical devices.
NOTE It is not a requirement of this part of ISO 17665 to have a complete quality management system during manufacture, but the elements of a quality management system that are the minimum necessary to control the sterilization process are normatively referenced at appropriate places in the text (see, in particular, Clause 4). Attention is drawn to the standards for quality management systems (see ISO 13485) that control all stages of production of medical devices, including the sterilization process. Regional and national regulations for the provision of medical devices might require implementation of a complete quality management system and the assessment of that system by a third party.
1.2.5 This part of ISO 17665 does not specify requirements for occupational safety associated with the design and operation of moist heat sterilization facilities.
NOTE Requirements for operational safety are specified in IEC 61010-2-040. Additionally, safety regulations exist in some countries.
Best Sellers From Food and Agriculture Sector
GSO 150-2:2013
Gulf Standard
Expiration dates for food products - Part 2 :
Voluntary expiration dates

BH GSO 150-2:2015
GSO 150-2:2013
Bahraini Standard
Expiration dates for food products - Part 2 :
Voluntary expiration dates


BH GSO 2055-1:2016
GSO 2055-1:2015
Bahraini Technical Regulation
HALAL FOOD - Part 1 : General Requirements


GSO 2055-1:2015
Gulf Technical Regulation
HALAL FOOD - Part 1 : General Requirements

Recently Published from Food and Agriculture Sector
GSO CXS 327:2024
CXS 327:2022
Gulf Technical Regulation
Cumin

GSO 2197:2024
Gulf Technical Regulation
Coffee and Coffee Products - Roasted Coffee Beans

GSO 194-1:2024
Gulf Technical Regulation
Wheat Flour

BH GSO 2034:2024
CODEX STAN 257R:2007
Bahraini Technical Regulation
REGIONAL STANDARD FOR CANNED HUMUS WITH TEHENA